PMS PROCEDURE (continued) NM2C
( ) 6.2 SELF-DIAGNOSTICS
6.2.1 Connect the pipeline supply or open the cylinders.
6.2.2 Turn the System Power switch to ON.
6.2.3 Verify that the following is displayed on the alarm CRT:
VIDEO PASS NARKOMED 2C
FIRMWARE PASS COPYRIGHT 1993, NAD, INC.
STATIC RAM PASS VERSION X.XX NM2C SW
TIMER PASS SOFTWARE ID. XXXX
A/D CONVERTER PASS
AUDIO - PRIMARY PASS
- BACKUP PASS
SERIAL I/O PASS
CLOCK PASS
NON-VOLATILE MEMORY PASS
FUNCTIONAL
( ) 6.2.4 Record the machine software version on the header of the checklist
form.
Rev. K
6-4
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS