PMS PROCEDURE (continued) NM2C
PM Certification Procedure for Narkomed 2B/2C/GS Anesthesia System
1. Use the PM Certification form for Narkomed 2B, 2C, GS Anesthesia Systems.
2. Completely fill in the header information.
3. If a TEC 6 DESFLURANE vaporizer is mounted to the machine, a label stating:
"WARNING THE ADMINISTRATION OF DESFLURANE ANESTHESIA MAY
REQUIRE FRESH GAS CONCENTRATIONS HIGHER THAN COMMONLY USED
WITH OTHER VOLATILE ANESTHETIC AGENTS. O2 FRESH GAS
CONCENTRATION OF LESS THAN 21% MAY BE OBTAINED WITH HIGH
VAPORIZER SETTINGS. CONTINUOUS MONITORING OF THE O2
CONCENTRATION IN THE BREATHING SYSTEM IS THEREFORE REQUIRED.
DRAEGER MEDICAL, INC. RECOMMENDS THE CONTINUOUS MONITORING OF
THE CONCENTRATION OF ANESTHETIC VAPORS IN THE BREATHING SYSTEM
DURING THE ADMINISTRATION OF INHALATION ANESTHESIA."(part #
4112737-001) Shall be attached to the vapor box immediately above the TEC 6
vaporizer.
4. Check all vapor 19 and 19.1 vaporizers for correct labeling. All vaporizers must have a
label stating "THE CONCENTRATION OUTPUT OF THIS VAPORIZER SHALL BE
VERIFIED AFTER IT HAS BEEN ATTACHED TO THE ANESTHESIA MACHINE"
(part # S010015). This label shall be attached to the rear of the vaporizer directly below
the mount.
5. All Key Index Safety Systems vaporizers, (K.I.S.S.) must have a label stating
"CAUTION:
AFTER FILLING HAS BEEN COMPLETED, REINSERT PLUG INTO
UPPER FILLERPORTAND TIGHTENLOCKINGSCREW" (part #4112520-001). This
label shall be attached to the vaporizer directly above the keyed filler. Place a check
mark at "K.I.S.S. LABEL" on the PM Certification form.
6. Perform the vapor concentration test on all Dräger vapor vaporizers in accordance with
SP00073. For every vaporizer tested, fill out a "VAPOR VAPORIZER CALIBRATION
CHECK" label (part # S010016). Information on this label shall include your signature,
type of agent, date tested, test results @ 1%, 2.5%, 4%, and a PASS or FAIL indication.
This label shall be attached to the upper right side of the vaporizer. If vaporizer fails
the concentration test, check "NO" in the "RECOMMENDED FOR USE" section on the
PM Certification form.
Place a "CAUTION
DO NOT USE" label (part # 4114327) on the vaporizer, and issue
a departmental alert. The TSR shall also seek permission from the equipment operator
to remove the failed vaporizer from the machine and install a replacement vaporizer or
an adapter block onto the mount. All nonfunctional Dräger vapor vaporizers must be
removed from service for machine to receive certification.