NM2C PMS PROCEDURE (continued)
6.8.1.12 Inspect the following: canisters and gaskets, dust cup and O-ring,
condition of soda lime.
6.8.1.13 Are the absorber canisters and dust cup in good condition? ___ (Y)
6.8.1.14 Is the cm H
2
O gauge at zero (0)? ___ (Y)
6.8.1.15 Remove the O
2
Med sensor plug from the inspiratory valve dome
adapter and examine the two O-rings at the bottom of the plug.
Rev. P
6-14A