EasyManua.ls Logo

Dräger Narkomed 2C - Page 165

Dräger Narkomed 2C
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NM2C PMS PROCEDURE (continued)
6.8.7.1.5 Turn the System Power switch to ON, turn up the oxygen
flow until the system pressurizes to 30 cmH
2
O.
( ) 6.8.7.1.6 What is the value indicated on the flowmeter? ___
(60ml/min).
6.8.7.1.7 Remove all test equipment, and turn the System Power
switch to STANDBY.
6.8.7.2 Inspiration Valve Leak
6.8.7.2.1 Connect a test terminal to the inspiratory valve.
6.8.7.2.2 Connect a tee fitting (P/N 4109292) and calibrated pressure
meter to the test terminal.
6.8.7.2.3 Connect a pressure bulb to the open port of the tee adapter.
6.8.7.2.4 Connect another test terminal to the bag connector.
6.8.7.2.5 Connect a Capnomed flowmeter to the test terminal on the
bag mount.
6.8.7.2.6 Pressurize the system to 30 cmH
2
O.
( ) 6.8.7.2.7 What is the value indicated on the flow meter? ___ (60
ml/min).
6.8.7.2.8 Remove all test equipment.
6.8.7.3 Flow Direction
6.8.7.3.1 Attach a breathing circuit with a 3-liter bag at the Y-piece to
the inspiration valve and expiration valve or the expiratory
hose terminal on the ultrasonic flow sensor, if applicable.
6.8.7.3.2 Attach a 3-liter bag to the swivel bag mount.
6.8.7.3.3 Turn the System Power switch to ON.
6.8.7.3.4 Set the O
2
flow to 4 l/min.
6.8.7.3.5 Inflate the simulated lung by briefly using the O
2
Flush.
6.8.7.3.6 Partially open the APL valve.
Rev. P
6-19
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS

Table of Contents

Related product manuals