PMS PROCEDURE (continued) NM2C
( ) 6.15.5 Carbon Dioxide Flowmeter - if applicable
* 6.15.5.1 Set the gas selector switch to ALL GASES.
6.15.5.2 Open the carbon dioxide gas cylinder valve.
6.15.5.3 Is it possible to adjust the flow of the carbon dioxide over its range
of 600 ml/min.? ___ (Y)
6.15.5.4 Close the carbon dioxide flow control valve.
6.15.5.5 Is the correct flow control knob and label attached to the carbon
dioxide flow control valve? ___ (Y)
*These items do not apply to machines with later design (P/N 4113329-001) ORC.