EasyManua.ls Logo

Dräger Narkomed 2C - Page 182

Dräger Narkomed 2C
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PMS PROCEDURE (continued) NM2C
( ) 6.15.5 Carbon Dioxide Flowmeter - if applicable
* 6.15.5.1 Set the gas selector switch to ALL GASES.
6.15.5.2 Open the carbon dioxide gas cylinder valve.
6.15.5.3 Is it possible to adjust the flow of the carbon dioxide over its range
of 600 ml/min.? ___ (Y)
6.15.5.4 Close the carbon dioxide flow control valve.
6.15.5.5 Is the correct flow control knob and label attached to the carbon
dioxide flow control valve? ___ (Y)
*These items do not apply to machines with later design (P/N 4113329-001) ORC.
Rev. K
6-36
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS

Table of Contents

Related product manuals