PMS PROCEDURE (continued) NM2C
6.16.16 Select the OXYGEN HIGH alarm limit. Does a box appear around the
high alarm limit? ___ (Y)
6.16.17 What is the high oxygen alarm default? ___ % (100)
6.16.18 Verify that the high alarm limit has a range from 100 to 19%.
6.16.19 Set the high alarm limit to 95.
6.16.20 The message INSP O2 HIGH shall appear on the remote display under
the Advisory column.
6.16.21 Return the high alarm limit to 100.
6.16.22 The INSP O2 HIGH message shall disappear.
( ) 6.16.23 Within 3 minutes, what is the oxygen concentration? ___ % (97-100)
Rev. K
6-38