NM2C PMS PROCEDURE (continued)
6.25A.20 Adjust the O
2
flow to 3 l/min.
( ) 6.25A.21 What is the tidal volume on the test volumeter? ___ml (>250)
6.25A.22 Remove the ventilator hose from the ventilator hose terminal.
6.25A.23 Attach a test terminal to the bellows assembly ventilator hose terminal.
6.25A.24 Connect a flowmeter test stand (P/N S000081) to the test terminal.
6.25A.25 Set the frequency to 1 BPM.
6.25A.26 Set the I:E RATIO to 1:1.
( ) 6.25A.27 What is the flow indicated during the inspiratory phase? ___(<50 ml)
6.25A.28 Remove the test terminal and flowmeter test stand.
6.25A.29 Return all controls to their original position.
Rev. Q
6-51
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS