PMS PROCEDURE (continued) NM2C
6.30 OXYGEN FLUSH VALVE
6.30.1 Press and release the O
2
FLUSH button.
6.30.2 Does the flow of oxygen stop immediately? __ (Y)
6.30.3 Close the APL valve.
6.30.4 Connect a 12-inch hose to the inspiratory valve.
6.30.5 Set the Man/Auto selector to BAG.
6.30.6 Occlude the bag mount.
6.30.7 Insert the sensor from a calibrated O
2
Med into the valve dome adapter
on the inspiratory valve.
6.30.8 Close all flow control valves.
6.30.9 Press the O
2
FLUSH button.
( ) 6.30.10 What is the O
2
concentration after the value stabilizes?___%O
2
(97-100)
6.30.11 Remove the O
2
Med sensor and install the plug.
6.30.12 Remove thetestminute volumeter (P/N 2212300) from the absorber and
connect it to the common gas outlet, using Fresh Gas Outlet Volume
Test Device (P/N S010158).
6.30.13 Press and hold the O
2
FLUSH button for 15 seconds; multiply the value
by 4.
( ) 6.30.14 What is the oxygen flush flow rate? ___l/min. (45-65)
6.30.15 Remove the test minute volumeter and test fixture, and reconnect the
fresh gas hose.
6.30.16 Turn the System Power switch to ON.
Rev. L