TROUBLESHOOTING GUIDE (continued) NM2C
GUIDE 3: Loss of Respiratory Volume Monitor
Y
Y
Y
N
N
N
Y
N
N
Y
N
Y
N
Y
START
REPLACE
PROCESSOR ASSEMBLY
AS OUTLINED IN
SECTION 4
REINSTALL ORIGINAL
PROCESSOR ASSEMBLY
CONTACT NAD
SERVICE DEPT.
UNIT IS
FUNCTIONAL
DOES UNIT
PASS PMS?
IS RESPIRATORY
VOLUME MONITOR
WORKING PROPERLY?
IS RESPIRATORY
VOLUME MONITOR
WORKING PROPERLY?
REINSTALL ORIGINAL
VOLUME SENSOR
---------------------------------
CHECK INTERNAL CABLE
FROM INTERFACE PANEL
TO J17 ON
PROCESSOR BOARD
PERFORM A
COMPLETE PMS
ON UNIT
IS RESPIRATORY
VOLUME MONITOR
WORKING PROPERLY?
REPLACE VOLUME SENSOR
AS OUTLINED IN
SECTION 4
CHECK SENSOR
AND VALVE
MOUNTING & GASKETS
CONNECT SENSOR
CORD TO
INTERFACE PANEL
IS
SENSOR INSTALLED
PROPERLY?
IS SENSOR
CORD CONNECTED TO
INTERFACE PANEL?
IS RESPIRATORY
VOLUME MONITOR
WORKING PROPERLY?
3-6
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS