12 Pocket Guide Oxylog 3000 plus
Assembly
Assembling the adult reusable hose
system
1 Place the diaphragm (B) in the breathing valve hou-
sing (C). Ensure that it is inserted correctly.
2 Fit the cover (A) and turn it approximately 60
o
clock-
wise to secure into position (a click can be felt).
3 Push the flow sensor (D) onto the breathing valve
(C). Note the correct alignment of the parts by the
groove in the flow sensor (D) and the notch on the
breathing valve (C).
4 Push the angled connector (E) onto the flow sensor
(D).
Hose connections
1 Connect the ventilation hose (F) to the breathing
valve.
2 Connect the flow measuring hoses (G) to the
nozzles on the flow sensor. Note the different dia-
meters of the hoses and the nozzles when connec-
ting the flow measuring hoses and connect to the
correct side.
3 Connect the flow measuring hoses (G) to the
Oxylog 3000 plus. Correct alignment is indicated by
a notch on the connector, which must point away
from the ventilation hose. Otherwise, the set will not
fit and the measured values will be incorrect.
4 Connect the ventilation hose (F) to the gas outlet on
the Oxylog 3000 plus.
When connecting a hose, check that the hose setting in
the Settings window corresponds to the connected ho-
se.
009011
A
B
C
D
E
F
G
G
012
F
G

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals