26 Pocket Guide Oxylog 3000 plus
Operation
Setting pressure support PC-BIPAP/PS
The following can also be set on the display for
PC-BIPAP:
Setting on page 1: Pressure support
Δ
Psupp above
PEEP.
Setting on page 2: Pressure rise time Slope.
Flat slope = long pressure rise time
Medium slope = medium pressure rise time
Steep slope = short pressure rise time.
SpnCPAP, SpnCPAP/PS
Spontaneous Continuous Positive Airway Pressure
Set the ventilation pattern with the controls below the
display:
Maximum airway pressure Pmax.
–O
2 concentration FiO2.
The following can be set on the display:
Positive end expiratory pressure PEEP.
Pressure support
Δ
Psupp above PEEP.
NIV - Non-invasive ventilation.
Insp.term.%PIF
Inspiration termination criterion of pressure suppor-
ted strokes, as percentage of the peak inspiratory
flow (PIF).
Hose type
The selected hose type must match the hose type in
use. Otherwise a correct volume measurement can-
not be guaranteed.
Cuvette type (optional)
If
Δ
Psupp is set above 0 mbar, the following can also be
set on the display for SpnCPAP:
Sensitivity Trigger.
Successful patient triggering is indicated by an aste-
risk (*) on the left side of the curves window.
Pressure rise time Slope (effective for pressure
support
Δ
Psupp).
12437171
15737171

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals