28 Pocket Guide Oxylog 3000 plus
Operation
Special functions
Manual inspiration / Inspiration hold
The function Manual inspiration / Inspiration hold will eit-
her initiate a new (manual) ventilation stroke or hold the
inspiratory phase of the current ventilation stroke for a
maximum of 15 seconds.
The pattern of the manually started ventilation stroke
corresponds with the set ventilation mode.
This function is not available for:
– SpnCPAP without PS,
–O
2 inhalation (optional).
To activate Manual inspiration or Inspiration hold
z Press key Inspiration hold (A) for as long as inspi-
ration is required.
100 % O2 (optional)
To apply 100 % O2 for 3 minutes regardless of the mo-
mentarily set value.
z Briefly press key 100 % O
2 (B).
Its indicator lights up for 3 minutes.
The set value is resumed by the ventilator upon expiry
of these 3 minutes, or when the 100 % O
2 is pressed
again. The indicator dims.
O2 inhalation (optional)
The O2 inhalation function is not a ventilation mode.
It may only be used for patients with spontaneous brea-
thing who receive a constant O
2 flow of between 0 and
15 L/min via a mask.
To activate O
2 inhalation
1 Connect the inhalation mask to the gas outlet for
ventilation hose (B).
2 Press and hold key O
2-Inhalation (A) for approx.
3 seconds.
O
2 inhalation is performed with the previously effec-
tive setting.
3 Set and confirm the required O
2 flow via the rotary
knob (C).
01237170
08337170
Oxylog 3000 plus
C
A B