Page #21 background imageLoading...
Page #21 background image
Pocket Guide Oxylog 3000 plus 21
Operation
Operation
Ventilation functions of the
Oxylog 3000 plus
Ventilation modes:
Volume-controlled ventilation:
VC-CMV / VC-AC
–VC-SIMV
Pressure-controlled ventilation:
–PC-BIPAP
Support of spontaneous breathing:
SpnCPAP
Additional settings for ventilation:
Pressure support: in the ventilation modes
VC-SIMV, PC-BIPAP, SpnCPAP
Apnoea ventilation: in the ventilation mode
SpnCPAP
AutoFlow (optional): in the ventilation modes
VC-CMV, VC-AC, and VC-SIMV
NIV: in the ventilation modes SpnCPAP (/PS),
PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF, and
VC-SIMV / AF
Special procedures:
Inspiration hold
–O
2 inhalation (optional), with an inhalation mask
VC-CMV, VC-AC
Volume Controlled - Controlled Mandatory
Ventilation
Volume-controlled ventilation with fixed mandatory mi-
nute volume MV, which is set with tidal volume VT and
respiratory rate RR.
Set the ventilation pattern with the controls below the
display:
Tidal volume VT.
Ventilation respiratory rate RR.
(minimum possible frequency: 5 per min).
Maximum airway pressure Pmax.
–O
2 concentration FiO2.
NOTE
The Oxylog 3000 plus Pocket Guide is not a replacement or substitute for the Instructions for Use.
Any use of the device requires full understanding and strict observation of the Instructions for Use.
503
Paw
Flow
T
insp
T
e
P
plat
Plateau time
Tplat
1
RR
Insp. Flow
t
t
P
max
PEEP
Exp. Flow

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals