Pocket Guide Oxylog 3000 plus 17
Getting started
Error messages during the device check
Message Cause Explanation/Remedy
System leakage Leak in ventilation hose system
and/or test lung.
Check hoses, breathin valve, flow
sensor, and test lung for leaks and
replace if necessary.
Internal leak in system. Contact your local DrägerService
for additional support.
No test lung Test lung not connected or major
leakage.
Connect test lung.
Check hoses, breathing valve, flow
sensor, and test lung for leaks and
replace if necessary.
Breathing valve inop Breathing valve has malfunctioned. Check correct condition of brea-
thing valve including diaphragm
and rubber disc; fit a new breathing
valve if necessary or use a new dis-
posable hose set.
Pressure measurement inop The ventilation hose system has not
been connected correctly.
Connect ventilation system cor-
rectly.
Pressure measurement is not pos-
sible.
Contact your local DrägerService
for additional support.
PEEP-valve inop Internal leak in system. Check hoses, breathing valve, flow
sensor, and test lung for leaks and
replace if necessary.
Device defective. Contact your local DrägerService
for additional support.
Patient flow measurement inop Flow measurement implausible. Replace flow sensor.
Contact your local DrägerService
for additional support.
Hose detection inop The device check failed on the hose
detection.
Connect a different hose or change
hose setting.
Detected hose differs from se-
lected hose
The hose that is detected differs
from the selected hose type, or the
flow measuring hoses are incorrect-
ly positioned.
Connect a different hose or change
hose setting.