EasyManuals Logo

Dräger Oxylog 2000 plus Pocket Guide

Dräger Oxylog 2000 plus
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
30 Pocket Guide Oxylog 3000 plus
Operation
5 Confirm. The zero calibration starts and the line dis-
plays Busy.
Note the possible warm-up time. During zero calib-
ration, ventilation settings can be changed. After a
successful zero calibration, the line briefly displays
Pass.
6 Attach the CO
2 sensor (A) back to the cuvette (B).
If zero calibration was not successful:
The Oxylog 3000 plus displays the alarm
!!! CO
2 Zero calib. failed.
z Repeat zero calibration
If zero calibration is still not possible:
1 Check whether the sensor (A) is soiled and clean if
necessary. If the sensor is defective, replace the
sensor.
2 Repeat zero calibration.
CO2 filter check during ventilation
To perform CO2 filter check:
1 Remove the CO
2 sensor from the cuvette.
2 Attach the CO
2 sensor (A) to the test filter (B).
3 Press the Settings key.
4 Select and activate the line CO2 Filter check - Run.
5 Confirm. The filter check starts and the line displays
Busy.
During the filter check, ventilation settings can be
changed.
After a successful filter check, the line briefly dis-
plays Pass.
6 Attach the CO
2 sensor (A) back to the cuvette.
If the check was not successful:
The Oxylog 3000 plus displays the alarm !!! CO2 Filter
check failed. The test value is outside the permissible
tolerance.
1 Check whether the sensor (A) or test filter (B) is
soiled and clean them if necessary. If the sensor is
defective, replace the sensor.
2 Check the CO
2 calibration with test gas or calibrate
the CO
2 sensor.
026
A
B

Other manuals for Dräger Oxylog 2000 plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Oxylog 2000 plus and is the answer not in the manual?

Dräger Oxylog 2000 plus Specifications

General IconGeneral
BrandDräger
ModelOxylog 2000 plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals