Cleaning, Disinfection and Sterilization
100
Instructions for Use Oxylog 3000 SW 1.n
Disinfecting by wiping
Ventilator and medical gas hose:
– Follow the manufacturer's instructions. Remove
heavy soiling with a disposable cloth first.
Bath disinfection for reusable hose system
Disassembled parts of the breathing valve, flow
sensor, ventilation hose and flow measuring hoses:
Sterilizing the reusable hose system
– Disassemble the breathing valve, flow sensor
and angled connector. When disassembling the
breathing valve from the flow sensor, pull in a
straight line.
Do not rotate the parts, as this may damage the
flow sensor. Dismantle the breathing valve.
WARNING
Do not allow any liquid to enter the ventilator
or medical gas hose!
Risk of malfunction.
WARNING
Follow the manufacturer's instructions.
Agitate parts thoroughly in the solution.
Do not clean with a hard brush!
WARNING
Do not allow any objects to enter the breathing
valve or flow sensor!
Risk of malfunction.
WARNING
Rinse parts thoroughly with distilled water.
Disinfectant residues can cause the rubber
disc to become jammed in the breathing
valve!
WARNING
Allow to dry completely. The breathing valve
and flow measuring hoses may not function
correctly if water remains in these parts!
Do not purge with compressed air. The
silicone diaphragm may be damaged or
misplaced. Malfunction of the breathing valve
can endanger the patient.
Risk of CO2 rebreathing.