EasyManua.ls Logo

Dräger Oxylog 3000 - Measured Value Display; Airway Pressure Measurement; Flow Measurement; Frequency Measurement

Dräger Oxylog 3000
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use Oxylog 3000 SW 1.n 115
Technical Data
Measured value display
Airway pressure measurement
Range 0 to 99 mbar
Resolution 1 mbar
Accuracy ±2 mbar
Maximum airway pressure Ppeak
Positive end expiratory pressure PEEP
Mean airway pressure Pmean
Plateau pressure Pplat
Flow measurement
Minute volume MV
Range 0 to 99 L/min, BTPS
Resolution 0.1 L/min
Accuracy ±15% of measured value, or ±1 L/min, whichever is
greater.
Tidal volume VTe
Range 0 to 5000 mL, BTPS
Resolution 1 mL
Accuracy ±15% of measured value, or ±25 mL, whichever is
greater.
Frequency measurement
Range 0 to 99 /min
Resolution 1 /min
Accuracy ±1 /min
Curve display
Airway pressure Paw (t) –10 to 100 mbar
Flow (t) –120 to 120 L/min