Technical Data
118
Instructions for Use Oxylog 3000 SW 1.n
AC/DC power pack
Temperature range –20
o
C to 50
o
C
Protection class to
EN 60601
Class II, the earthing is used for EMC purposes.
Input 100 to 240 V~/ 50 to 60 Hz / 0.6-0.3 A
Output 19 V ±0.5 V / 2.1 A
To isolate the ventilator system from mains,
disconnect the power cord from the wall connector.
DC/DC converter
Temperature range –20
o
C to 50
o
C
Input 10 / 24 / 28 V DC; 9 A / 4.5 A / 4 A
Output 19 V ±0.5 V / 2.1 A
Gas supply
From a pipeline system or from a medical gas
cylinder.
O
2 service pressure 300 kPa –10 % to 600 kPa at 100 L/min
Supply gas Medical oxygen or medical air.
O
2 inlet connection either:
NIST
1)
to EN 739, or
DISS
2)
to CGA V5-1989, or
N-F
3)
S90-116
WARNING
Only use medical grade oxygen or medical grade
air that is dry and free from dust and oil.
Contaminated gas can cause lung damage and
device malfunction.
Gas cylinders and
pressure reducers
WARNING
Only use compressed gas cylinders and
pressure reducers, which comply with all
applicable regulations and have been approved.
Pressure reducer Must have a vent valve on the output side to limit the
delivery pressure to approximately 1000 kPa in the
event of a fault.
Gas consumption for
internal control
Average 0.5 L/min
Accuracy of gas
consumption indication
15 % or ±1 L/min, whichever is greater.
1) NIST = Non Interchangeable Screw Thread Connection
2) DISS = Diameter Index Safety Systems
3) N-F = French standard