Instructions for Use Oxylog 3000 SW 1.n 113
Technical Data
Settings
Ventilation modes IPPV (CMV)/IPPVAssist (CMVAssist), SIMV,
SIMV/ASB (SIMV/PS), BIPAP
®
(PCV+),
BIPAP
®
/ASB (PCV+/PS), CPAP, CPAP/ASB
(CPAP/PS)
Ventilation frequency Freq. 2 to 60 1/min ±1 1/min (SIMV, BIPAP
®
)
5 to 60 1/min ±1 1/min (IPPV, IPPVAssist)
12 to 60 1/min ±1 1/min for apnoea ventilation
Ventilation time ratio I:E (IPPV, IPPVAssist) 1:4 to 3:1
Inspiration time Tinsp (SIMV, SIMV/ASB, BIPAP
®
,
BIPAP
®
/ASB)
0.2 to 10 s
Tidal volume VT 0.05 to 2.0 L, BTPS
1)
Accuracy of Setting ±15% of set value or ±25 mL, whichever is greater.
Inspiratory pressure Pinsp PEEP+3 mbar to 55 mbar
2)
O2 concentration 40 to 100 vol.%
Accuracy of Setting
±10 vol.% The setting depends on the inspiratory
flow
3)
and mean airway pressure.
Positive end expiratory pressure PEEP 0 to 20 mbar ±2 mbar, no negative pressure.
Trigger sensitivity (flow trigger) 3 to 15 L/min
Pressure support ΔASB 0 to 35 mbar (relative to PEEP) ±2 mbar
Rise time for pressure
support
slow (1 s), standard (0.4 s), fast (0 s).
1) BTPS
Body Temperature, Pressure, Saturated.
Measured values referred to the conditions of the patient's lungs,
body temperature 37
o
C, airway pressure, water-vapor-saturated gas.
2) 1 mbar = 100 Pa
3) see O2 concentration, page 59