EasyManua.ls Logo

Dräger Oxylog 3000 - Connect the Test Lung; Perform Device Check; Switch on

Dräger Oxylog 3000
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use Oxylog 3000 SW 1.n 41
Getting Started
Perform device check
The device check consists of the following steps.
Duration is approximately 3 minutes.
Connect the test lung
1 Connect the angled connector (A) to the
breathing valve.
2 Connect the catheter connector (B), diameter
7 mm, to the angled connector.
The catheter connector simulates the
resistance of the airways.
3 Connect the test lung (C).
Switch ON
1 To switch ON briefly press the O key (A).
The device performs a self-test and the operator is
prompted, on the display, to activate the configura-
tion menu or device check:
Press rotary knob for device check and
configuration
2 Press the rotary knob (B) to confirm, before the
bar is full.
3 Select Device check in the main menu and
confirm.
021
CAUTION
BTPS values of a test lung are not the same as the
BTPS values of a patient. The Oxylog 3000 mea-
sures and adapts according to BTPS values of a
patient. Therefore, when a test lung is connected,
the MVe and VTe indicated on display may differ
from the MVe and VTe that is set by the
operator.
A
C
B
02211937171
NOTE:
The device check can be discontinued at any time
by pressing the Alarm Reset key.
Oxylog 3000
AB

Table of Contents

Related product manuals