Monitoring
Expiratory minute volume MV
Alarm, upper alarm limit When the upper alarm limit has been exceeded.
Range of settings 2 to 41 L/min
Alarm, lower alarm limit When the level drops below the lower alarm limit.
Range of settings 0.5 to 40 L/min
Airway pressure Paw
Alarm, upper alarm limit When value "Pmax" is exceeded.
Range of settings 20 to 60 mbar
Alarm, lower alarm limit When the pressure difference between inspiratory
and expiratory phases is less than 5 mbar.
Or
If the set pressure level is not attained.
Apnoea alarm time TApnoea
Alarm When respiratory activity is no longer detected.
Range of settings 15 to 60 s, can be set in 1 s increments.