Operation
54
Instructions for Use Oxylog 3000 SW 1.n
Pressure support
The following can also be set on the display for
BIPAP
®
, BIPAP
®
/ ASB:
Setting on page 1: Pressure support
Δ
ASB
above PEEP.
Setting on page 2: Pressure rise time Ramp
flat ramp = long pressure rise time
medium ramp = medium pressure rise time
steep ramp = short pressure rise time.
–BIPAP
®
, BIPAP
®
/ASB can be extended to
include the application mode NIV – Non-inva-
sive ventilation, see page 58.