Getting Started
40
Instructions for Use Oxylog 3000 SW 1.n
Checking readiness for operation
Whenever the ventilator has been serviced or
the ventilation hoses changed.
At the latest every six months.
The following functions are checked with the menu-
based test:
Gas supply present.
Hose system / breathing valve connected and
OK.
Alarm functions OK.
Ventilation functions OK.
Monitor functions OK.
Oxylog 3000 interrupts the test if a fault is detected.
The relevant fault is indicated on the screen.
WARNING
The patient may be endangered if the above
device check is not completed.