Instructions for Use Oxylog 3000 SW 1.n 29
Assembly
Assemble the reusable hose system
Parts must always be sterilized before use!
Breathing valve assembly
1 Place the diaphragm (B) in the breathing valve.
Ensure that it is inserted correctly.
2 Fit the cover (A) and turn it approximately 90
o
clockwise to secure into position.
3 Push the flow sensor (C) into breathing valve.
Note the preferred position as indicated by the
groove.
4 Push the angled connector (D) onto the flow
sensor.
008
WARNING
The rubber disc (A) in the housing may not be
removed, damaged or bent, otherwise the
valve will not work properly and will endanger
the patient.
Risk of CO2 rebreathing.
A
009
WARNING
Always use an angled connector. If the angled
connector is not used, the minute volume may
be measured incorrectly.
A
B
C
D