Maintenance
104
Instructions for Use Oxylog 3000 SW 1.n
Maintenance intervals
Inspection examination of actual condition.
Service measures to maintain specified
condition.
Repair measures to restore specified
condition.
Maintenance inspection, service, and repair,
where necessary.
Preventive
Maintenance
maintenance measures at regular
intervals.
CAUTION
In order to avoid malfunctioning of the device,
maintenance must be carried out by properly
trained service personnel.
CAUTION
Have the ventilator inspected and serviced at
regular two-year intervals.
Keep a record on all preventive maintenance.
WARNING
Clean and disinfect the device or device parts
before each maintenance step – and also
when returning for repair to prevent risk of
infection.
Task Frequency Performed by
Replace dust filter.
1)
Every two years. Trained service personnel.
Replace internal battery. Every two years
When the battery no longer
remains charged for the speci-
fied operating time
2)
.
Trained service personnel.
Device inspection and maintenance. Every two years. Trained service personnel.
1) The dust filter can be treated as household waste.
2) Refer to "Technical Data" section for the battery operating time.