Operation
60
Instructions for Use Oxylog 3000 SW 1.n
When patients are breathing spontaneously, the
achievable O2 concentration will depend on the
profile of the inspiratory flow. Even if this profile is
changed, the message "Check settings O
2" may
appear after some time if the desired concentration
can not be reached.
WARNING
In toxic surroundings:
The patient must be ventilated with 100%
O
2 so that toxic constituents do not enter
into the breathing gas.
The patient must be immediately trans-
ferred to a breathable atmosphere in order
to prevent inhalation of toxic air when
spontaneous breathing resumes.