EasyManuals Logo

Dräger Polytron 7000 User Manual

Dräger Polytron 7000
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #119 background imageLoading...
Page #119 background image
119
Metrological certificate of approval
Information on DrägerSensor O
2
(6809720)
Principle of measurement: The DrägerSensor O
2
(6809720) is an electrochemical two-electrode-sensor for measurement of oxy-
gen (O
2
) in air. The sensor can only be operated in connection with a suitable Dräger transmitter.
Reaction at the measuring electrode: O
2
+ 2 H
2
O + 4 e
=> 4 OH
Reaction at the counter-electrode: 2 Pb => 2 Pb
2
+ + 4 e
Ambient conditions:
Operational Characteristics –5
o
C to 40
o
C, temporarily –20
o
C to 55
o
C
700 hPa to 1300 hPa
10 to 95 % relative humidity, non condensing
Storage 0
o
C to 40
o
C
700 hPa to 1300 hPa
30 to 70 % relative humidity, non condensing
Measured value time setting:
t
0...20
<12 seconds
t
0...90
<30 seconds
Note: In the case of temperatures below –5
o
C, the measured value time setting may increase.
Calibration:
Flow rate: 0.5 L/min
Zero gas Nitrogen (99.9 Vol.-% N
2
)
Calibration gas Oxygen / nitrogen – mixed gas
Warming-up time:
Sensor is ready to operate after 15 minutes
Sensor is ready to calibrate after 2 hours

Table of Contents

Other manuals for Dräger Polytron 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 7000 and is the answer not in the manual?

Dräger Polytron 7000 Specifications

General IconGeneral
BrandDräger
ModelPolytron 7000
CategoryAdapter
LanguageEnglish