Start-up
38
Start-up
Switch on power supply.
The transmitter begins its warm-up routine:
Depending on the type of sensor installed, the warming-up period time may
last between 5 minutes and 12 hours. See the related information in the op-
erating instructions for the sensor. The warm-up phase may take longer in
extremely high or low temperatures.
When the sensor has warmed up:
Calibrate sensor, page 44, when a pre-calibrated sensor is not used.
Transmitter is ready for use.
Check the transmission of the signals between to the control unit and the
initiation of alarms, see the function group » Analogue interface « on
page 71.
The software version, the date and the time are displayed.
NOTE
For the correct operation and functionality it is important to set the date
and time.
The sensor is now warming up. The remaining warming-up time for the
sensor is displayed.
NOTE
Note on the operation with relay module:
During the warm-up period, the relay module indicates a fault. Alarms A1
and A2 are not indicated during the warm-up period.
The sensor is ready for use.
Pioneering Solutions
Polytron 7000
A001_en.eps
A003_en.eps
SWversion: 8.0
29.10.11 11:47
A004_de.eps
H
2
S
ppm
in 00
:
01
:
30
Sensor ready
A005_en.eps
H
2
S
ppm
0.0

Table of Contents

Other manuals for Dräger Polytron 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 7000 and is the answer not in the manual?

Dräger Polytron 7000 Specifications

General IconGeneral
BrandDräger
ModelPolytron 7000
CategoryAdapter
LanguageEnglish