EasyManuals Logo

Dräger Polytron 7000 User Manual

Dräger Polytron 7000
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
Fault – Cause – Remedy
48
Fault – Cause – Remedy
If the display will not function: Have the transmitter checked by
DrägerService.
The fault and warning numbers shown in the following tables are displayed in
the menu under » Information «, » Instrument «, » Fault « or » Warnings « –
see page 55.
Fault number Cause Remedy
#1 Serious data error in unit
various causes.
Initialise the unit with the menu items » Settings «,
» Instrument «, » Init. device «, page 68.
If this error occurs again: have the transmitter checked
by DrägerService.
#2 Serious unit fault –
various causes.
Have the transmitter checked by DrägerService.
#61 Data error on the interface card –
various causes
Have the transmitter checked by DrägerService.
#63 Hardware fault or software error in the
pump module.
Change the pump module, page 34.
#64 Gas flow of the pump module falls
below the configured fault threshold.
Reliable measurements are no longer
possible.
Check the hoses for blockages,
If necessary adjust pump flow.
#65 Open-circuit in 3-wire cable. Check the connections.
#67 Bad contact of the relay module. Check connector of the relay module or fit it again.
# 100 Unit cannot detect a sensor. Remove the sensor and install it again, page 46.
If the problem persists, check the sensor plug or install
a new sensor.
# 101 Sensor data error in the unit. Remove the sensor and install it again, page 46.
If this error occurs again: have the transmitter checked
by DrägerService.
# 102 Unit does not support this
sensor version.
Use a compatible sensor, see the ordering list on
page 93.
# 103 Sensor data error in the unit. Initialise the sensor with the menu items » Settings «,
» Sensor «, » Sensor-EC «, » Init. sensor «, page 78.
If this error occurs again: have the transmitter checked
by DrägerService.
# 106 Zero-point not correctly. Zero-point calibration, page 58.
# 107 Sensitivity calibration not executed cor-
rectly.
Repeat sensitivity calibration, page 59.
# 108 Sensor data error. Replace the sensor, page 46.
# 109 Unit fault. Check the sensor contacts; otherwise have the transmit-
ter checked by DrägerService.
# 121 Fresh-air calibration (first step of auto-
calibration) not executed correctly.
Repeat autocalibration, page 60. Make sure that the
ambient air is free of other gases.
# 125 Sensor not ready. Replace the sensor, page 46.
# 129 Electrolyte liquid evaporates Refill electrolyte. See sensor data sheet.

Table of Contents

Other manuals for Dräger Polytron 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 7000 and is the answer not in the manual?

Dräger Polytron 7000 Specifications

General IconGeneral
BrandDräger
ModelPolytron 7000
CategoryAdapter
LanguageEnglish