Maintenance
Calibration with test-gas ampoules
Use the calibration flask (Part No. 68 03 407) – Observe the information in
the sensor data sheet.
Follow the instructions printed on the calibration flask and enclosed with the
test-gas ampoules.
For units without a pump:
1 Fit the adapter and calibration bottle to the Polytron 7000.
Break the test-gas ampoule inside the calibration flask.
Wait until the measured value has settled (see the operating instructions for
the sensor for the necessary waiting period). Then carry out calibration.
For units with a pump:
The pump adapter is already installed on the Polytron 7000.
Insert the adapter (Part No. 68 04 620) in the calibration flask – connect the
hoses.
Break the test-gas ampoule inside the calibration flask and wait until the
measured value has settled (see the operating instructions for the sensor
for the necessary waiting period). Then carry out calibration.
05023758_1.eps
OK
68 07 407
04923758_1.eps
OK
+
05123758_1.eps
OK
68 07 407
68 04 620
+

Table of Contents

Other manuals for Dräger Polytron 7000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Polytron 7000 and is the answer not in the manual?

Dräger Polytron 7000 Specifications

General IconGeneral
BrandDräger
ModelPolytron 7000
CategoryAdapter
LanguageEnglish