The menu » Calibration «
The required value and the actual value are now displayed.
Disconnect the flow of calibration gas.
If these are correct as displayed:
Select » Accept value « and press the » « key.
CAUTION
An alarm could be triggered in the central unit for as long as the calibration
gas concentration is pending!
OK