Dymax BlueWave® 200 Version 3.0 User Guide
Frequently Asked Questions
Q. My BlueWave 200 will not turn on.
Check the Power Cord connection.
Check the Fuses located where the Power Cord plugs into the BlueWave 200.
Q. The Bulb will not ignite; it only “flickers”.
Replace the Bulb. Excessive power cycling will shorten the life expectancy of the Bulb. This is an arc,
not a filament Bulb. Once ignited, it must be left on for a minimum of 10 minutes to fully vaporize
elements in the Bulb. If not, the Bulb may be difficult to re-ignite. Each re-ignition increases the rate of
Bulb degradation.
Q. I installed a new Bulb, and it still will not ignite.
Check to make sure the Bulb Connector is fully seated into the Igniter.
Q. Why do I have low intensity, even with a new Bulb?
The standard BlueWave 200 has a Filter Lens installed, which filters the light before it reaches the Light
Guide. The light intensity will be decreased as the light that passes through the filter is restricted by
dust and debris.
The intensity is being checked too early. The 200 Watt Bulb will not reach full intensity until five
minutes after initial power up.
The Lightguide may not be fully seated into the Lightguide Mount.
The end of the Lightguide may have a build-up of adhesive. Carefully remove with isopropyl alcohol or
use a razor blade for heavier deposits.
The condition of the Lightguide will also affect the intensity. All Lightguides degrade with time, but
intensity will also drop if the Lightguide is bent or compressed. The intensity reading from the
Lightguide should be compared to the intensity reading from a Lightguide Simulator to determine the
efficiency.
Incorrect installation of the Bulb. Bulb orientation is vital during the installation of the Bulb. Refer to
the Bulb installation instructions supplied with all new 200 Watt Bulbs for the proper installation
procedure.
Q. My Foot Pedal is not operating.
Check the connection of the Foot Pedal into the BlueWave 200.
Q. What causes my Shutter to hesitate to open when I activate my Foot Pedal?
Maintaining clean equipment and a clean working environment will help prevent the buildup of dust
and other debris from collecting on the mechanical parts of the BlueWave 200. Debris that settles in
the piston well (adjustment screw) can hinder the movement of the Shutter Solenoid Piston.
Shutter alignment to Reflector Mount incorrect.
The Shutter is a mechanical part which may wear after extended use.
Incorrect alignment of the Shutter to the Bulb Mount Assembly.