ELECTROMAGNETIC EMISSIONS AND IMMUNITY
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017 143
Table 2
Guidance and Manufacturers Declaration - Electromagnetic Immunity
The Dynatron Solaris Plus (and accessories) is intended for use in the electromagnetic environment
specied below. The customer or the user of the Dynatron Solaris Plus (and accessories) should assure
that it is used in such an environment.
Emissions Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
Compliant
Floors should be wood, concrete or
ceramic tile. If oors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
+/- 2 kV for power
supplyline
+/- 1 kV input/output lines
Compliant
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+/- 1 kV differential mode
+/- 2 kV common mode
Compliant
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % Ut
(>95 % dip in Ut)
for 0,5 cycle
40 % Ut
(60 % dip in Ut)
for 5 cycles
70 % Ut
(30 % dip in Ut)
for 25 cycles
<5 % Ut
(>95 % dip in Ut)
for 5 seconds
Compliant
Mains power quality should be that
of a typical commercial or hospi-
tal environment. If the user of the
equipment requires continued opera-
tion during power mains interrup-
tions, it is recommended that the
equipment be powered from an unin-
terruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m
Not appli-
cable
Power frequency magnetic elds
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.