COMBINATION THERAPY INSTRUCTIONS
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017 91
Combination
Therapy Instructions
e following combination therapy instructions are for Dynatron Solaris® Plus 708 and 709 USERS ONLY. e Dynatron
Solaris® Plus 705, 706, and 707 do not oer the Ultrasound feature.
Comboplus
™
Dynatronics’ Comboplus feature means you have almost unlimited options in setting up a combination treatment with
Solaris Plus. You can:
WARNING
• DO NOT use combination therapy for underwater treatment. Placing active electrodes underwater poses a
serious hazard to the patient!
• Use VERY LOW STIM INTENSITY for COMBO treatments.
• Remember to observe all contraindications, warnings, precautions, and usage cautions for BOTH Ultrasound
and Electrical Stimulation therapy when performing combination therapy.
• Since electrical current travels between the electrode and the soundhead during a COMBO treatment, the
electrode should be placed in proximity with the treatment area. Do not place the electrode and soundhead in
positions that will cause current to pass through contraindicated areas.
• Avoid removing the soundhead from the skin surface during “Stim rough Soundhead” treatments as
this may cause a momentary interruption of Stim current which may be uncomfortable to the patient. e
soundhead should remain in full contact of the skin until current output is stopped.
• Be alert for any sign of periosteal (bone) pain.