ULTRASOUND PROBLEM SOLVING
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017 87
If this error occurs, check to be sure the soundhead is rmly plugged into its connector. If you are unable to clear the message
by reconnecting the soundhead, contact Dynatronics’ customer service department at 1-800-874-6251 for assistance.
Miscellaneous
Certain conditions can cause an error in operation. When this occurs, the machine will not allow a treatment to be set up or
delivered and will display an error message. Some errors are easily resolved by the following methods.
Press STOP to stop the treatment, and turn the machine OFF then ON again. Always wait 5-10 seconds before
restarting the device.
Check to be sure the soundhead has not become disconnected from the machine. e soundhead should be rmly
plugged into its port. Only Dynatronics soundheads may be used with this device. If the soundhead has been
dropped, it may be damaged. If the device operates normally with one soundhead, but not with another, the problem
may be a damaged soundhead and you must contact Dynatronics Customer Service.
Make sure the soundhead is not too hot. In this case the Soundhead alert will appear in the Treatment Display Screen.
Check to see if conditions may have caused extreme moisture condensation in the device. is could occur when
the machine has become very cold then is brought indoors to a warm, humid environment. Condensation is a not
a serious condition. Allow the machine to sit in a dry environment until the condensation dries. e machine will
operate normally once the condensation is gone.
If you have tried all of these suggestions, the device may require service by the manufacturer. In this case, make a note of the
error message and the sequence of events that cause the error, and contact Dynatronics Customer Service at 1-800-874-6251
for further assistance. Do not send the device to Dynatronics without rst contacting the Customer Service Department.
Ultrasound Specications
Ultrasound Power output:
2cm
2
head: .........................1 MHz, 2 MHz, 3 MHz .................... 0-4 watts; 0-2.0 W/cm
2
± 10%
5cm
2
head: .........................1 MHz, 2 MHz, 3 MHz ..................0-10 watts; 0-2.0 W/cm
2
± 10%
10cm
2
head: .............................1 MHz, 2 MHz .........................0-20 watts; 0-2.0 W/cm
2
± 10%
10cm
2
head: ....................................3 MHz .................................0-10 watts; 0-1.0 W/cm
2
± 10%
Ultrasound Regulation and Technical Information
For the Dynatron Solaris® Plus 708, and 709 Only
e Dynatron Solaris Plus 708, and 709 comply with the following: