DIRECT CURRENT MODALITY INFORMATION
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017 62
Direct Current Warnings
It is important that excellent coupling be maintained any time current is owing. Increased resistance from poor coupling or
excessive current density can cause skin reactions.
• Electrodes must be attached and probe placed in contact with the patient’s skin prior to starting the treatment.
• Direct Current tends to lower surface sensory perception as the treatment proceeds. e patient must have good
sensory perception if Direct Current is to be used.
Direct Current Default Setting
Pulse Duration: ................................................................................... 100 mSec
Polarity: ..................................................................................................Negative
Intensity: .............................................................................................................1
e following default settings are set by the manufacturer:
Waveform: .................................................................................................Square
Pulse Duration: .............................................................................. .1-500 mSec
Pulse Duration Selections: .................................. 0.1, 0.3, 0.5, 1, 10, 100, 500
Power Output: ........................................................................................ 20 mA.
Note: Load impedance does not aect output of the DC waveform.