HIGH VOLT MODALITY INFORMATION
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017 49
Custom Contraction/Rest Time Cycles. e Dynatron Solaris Plus allows for the choice of muscle contraction and
relaxation time cycles (Duty Cycles) from options of 10/10 (ten seconds on and ten seconds o), 10/30, 10/50, Continuous
or Custom cycles. e Custom time cycle allows for a Contraction (ON) time from 1 to 120 seconds, and a Rest (OFF) time
from 1 to 300 seconds. Remember, the Rest time cannot be less than the Contraction time.
Selectable Ramp Speed. You can choose a ramp speed of .5 (1/2) to 2.0 seconds in half-second increments. For patient
comfort, the ramp occurs both before and aer the “Contraction” segment of the pulse.
Pulse Duration. e pulse duration is xed at 25 µs (micro-seconds).
Pulse Pair Interval. e interval between the two pulses in the wave form is xed at 100 µs.
Intensity Display in Volts. Intensity is displayed in volts (peak voltage with no load) with a range of 1 to 500 volts.
High Volt Default Settings
e following default settings are set by the manufacturer and are selected when you choose High Volt. You may change
these defaults to your own preferred settings. See “Setting Defaults” in this manual.
High Volt Pads Treatment
Continuous Duty Cycle
Treatment Time Electrodes: 10 minutes
Treatment Time Probe: 0.00
Polarity: Negative
Pulse Rate: High Range
Default High Range: ...........................................................................80-120 Hz
Default Low Range: ..................................................................................1-10 H
Available Range:....................................................................................1-200 Hz
High Volt Waveform Specications
Waveform: ................................................................... Twin peak, monophasic
Pulse Duration: ............................................................................................25 µs
Pulse Rate Range: .............................................................................1 to 200 Hz
Pulse Interval: ............................................................................................100 µs
Maximum Power Output: .................. Limited to less than 500 V, open load