INTERFERENTIAL / PREMODULATED INSTRUCTIONS
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017 28
• FIXED. A FIXED treatment automatically defaults to a STATIC setting. Press the toggle key under MODE until
FREQ is illuminated. A FIXED treatment can be set between 4,000 and 10,000 KHz. Selections progress from
4,000 KHz to 10,000 KHz in 1,000 KHz increments. Once the FIXED rate is set, return to the TIME display by
using the MODE toggle key.
e default High and Low frequency settings may be changed for a single treatment if desired, or new default settings
may be saved to apply to all future treatment setups.
• Press the toggle key under the MODE display. Select FREQ (Frequency).
• Press the toggle key under the RANGE display. Select a HIGH or LOW frequency range.
• e HIGH Frequency will appear on the le-hand side of the Treatment Display Screen. e LOW Frequency
will appear on the right-hand side of the Treatment Display Screen. Changes to the upper and lower limits are
made using the arrow keys to the side of each displayed setting. If you set both displays to the same value, the
treatment will be delivered at that single frequency rather than sweep through a frequency range.
• Aer pressing START, frequency settings will remain in eect for the duration of the treatment. If you save defaults
during this treatment, the new frequency settings you have entered become the defaults for this modality. However,
if you do not save the new settings, the unit will return to the current default settings for the next treatment.
• To SAVE these CUSTOM FREQUENCY settings, hold the START key down until a beep sounds. Once saved,
these settings will be used on all IFC or Premod treatments that follow.
• Press the toggle key under the MODE display to return to the TIME display, Aer 10 seconds with no key
presses, the Time display will automatically return.
4. Choose TARGET, SWEEP, or STATIC (for Interferential only)
Using the toggle key located under the TARGET display make your selection from the three available options.
TARGET. Pinpoint the treatment site delivering the full Interferential current where it is needed.
NOTE: Make any desired changes to the treatment time before selecting Consecutive. Treatment time changes
made aer selecting Consecutive will cause the treatment to revert to an ALTERNATING HIGH/LOW treatment.
NOTE: If INTENSITY has been set during the setup of a FIXED treatment option and a change is made to any
other treatment option, INTENSITY will automatically return to “0.”
NOTE: e intensity must be set before using the Target Pad so the patient will be able to indicate when the
treatment site is found. Also remember, an injured area will oen be more sensitive to the current delivered.
erefore, an intensity setting that is comfortable to the patient at rst may feel uncomfortable when the treatment
site is found using the TARGET PAD. If necessary, reduce the intensity to the patient’s comfort level.