HIGH VOLT INSTRUCTIONS
DYNATRON SOLARIS
®
PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017 47
8. Press START.
Electrodes Pad Treatment: Press START. When you press start, the treatment timer begins counting down.
Probe: Press START on the device followed by ON/OFF on the probe handle to activate a probe treatment. e timer
will begin counting up. If you fail to set the intensity on either an electrode or a probe treatment before pressing
START, the Intensity display will begin ashing, and you will be unable to start the treatment until the intensity is set.
9. MODIFY Settings
While the High Volt treatment is in progress, TIME, INTENSITY, AND POLARITY can be modied excluding
POLARITY for MultiStim probe. Carefully observe the channel indicator lights when modifying a treatment. When
a channel’s light is illuminated GREEN, the current treatment parameters for that channel are displayed. Any changes
made to the parameters will aect only the channel that is illuminated in GREEN. Use the Channels Toggle key to
display the parameters of another channel in order to modify parameters.
10. STOP
When the treatment time has elapsed, the current to the patient stops and a tone sounds signaling the end of a
treatment. Treatments in progress may be stopped at any time using one of the following methods.
Stop One Treatment Only. Press and hold the FUNCTION key and press STOP. is stops only the treatment in
focus.
Stop All. Press the STOP key. All treatments at all channels will stop.
Stop Time. Reduce the treatment time using the Time arrow key. e output at the selected channel is stopped,
and the device then displays the parameters for the next treatment that remains in progress (if there is an active
treatment).
Multistim Probe Stop. A probe treatment may be stopped by pressing the ON/OFF key located on the handle of
the MultiStim probe.
NOTE: e intensity arrows on the probe handle work the same as the intensity arrows on the device faceplate.
Never use High Volt to treat any conditions which contribute to loss of sensation, or an area where the patient
cannot feel the electrical stimulation.
NOTE: SAVING DEFAULTS. If the treatment you have just set up is the most common High Volt setup you use,
the treatment parameters can be saved as the defaults for your own device. Aer setting up the treatment, press
and hold the START key for two seconds. At the end of two seconds, a beep will sound indicating the treatment
parameters have been saved. e next time this modality is chosen, the parameters will be selected automatically.