XS series operating manual
XS user manual v0.55 - 14 -
1 µA current resolution (OPTION 116-00)
Add of a new range of current measure between 0 and 999 µA.
Value display
From 0 a 999 µA, display in µA
over 999 µA, display in mA
Threshold set up
With or without the option, the ΔI threshold can be set from 1 mA ±10 % to 10 mA ± 10% by 1 mA
step.
IMAX, IMIN and FIMAX threshold can be set from:
0 to 999 µA by 1µA step
1mA to 9.99mA by 0.01mA step.
Current measure
- In AC :
±(2.5% + 3 U) U = 1µA for measures between 0 and 200µA
±(2% + 2 U) U = 1µA for measures between 200 µA and 1mA
- In DC :
±(2% + 2 U) U = 1µA
Display of the value on a 999 points digital indicator.
In VDC accuracy is given for resistance loads > 1 M.