XS series operating manual
XS user manual v0.55 - 18 -
2 DIELECTRIC STRENGTH TEST 10VA
XS16 (10kVDC)
Output voltage
Continuous voltages only
From 100VDC to 10kVDC in one range.
Stability < 3% for V mains of 10%.
Negative pole to earth with DC voltage
Ripple < 0.1% for Is 1 mA
Accuracy of the output voltage : (2%+20volts) in relation with the set value for voltages
between 100 and 10000 volts and for a current < 100µA.
Discharge of the tested specimen and of the internal capacitor through a 1.5 M resistance
Voltage reading
By kilovoltmeter directly connected to the output terminals.
Accuracy : (1%+10 volts) of the read value.
Display by 2000 points digital indicator.
Short circuit current
≤ 1 mA for voltages between 100V and 10kVDC.
Breakdown detection
Current variation mode : I
The TEST detector automatically carries out the subtraction between the current which flows
normally in the device under test (I=U/Z) and the one which occurs at the time of a fault (I'=I+Ifault).
Amplitude fixed to 1 mA 10%.
Width of current pulse : 10µS 20%