KOSMOS User Guide 113
Environmental operating and storage conditions
Operating, charging, transport, and storage condition ranges
Mode of operation
Kosmos Bridge enforces scanning limits when using Kosmos Torso to maintain
safe user enclosurecontact temperatures.
Power supply (charger)
Rated input: 100-240V~, 50-60Hz, 1.5A
Watts: 60
Volts out: 5V, 5.8V, 8.9V, 11.9V, 15V, 20V
Current out (Amps): 4.6A, 4.6A, 4.4A, 4A, 3.6A, 3A
Operating Transport/Storage
Temperature (°C) 0C to +40C -20 C to +60 C
Relative humidity
(non-condensing)
15% to 95% 15% to 95%
Pressure 62 kPa to 106 kPa 62 kPa to 106 kPa
After storage at extreme temperatures, check the Kosmos Torso surface
temperature before applying to a patient. A cold or hot surface may burn
the patient.
Only operate, charge, and store Kosmos Bridge and Kosmos Torso within
the approved environmental parameters.
When used in high ambient temperatures (such as 40 deg C), the
KOSMOS safety feature may disable scanning to maintain safe touch
temperature.