KOSMOS User Guide 69
CHAPTER 7 Kosmos Torso
Kosmos Torso sheaths
Where fluid contamination is possible, cover Kosmos Torso with an appropriate
sterile sheath from CIVCO, which will promote asepsis and minimize cleaning.
Be aware that some patients have a latex allergy. Some commercially
available Kosmos Torso covers contain latex.
To prevent cross-contamination, use sterile transducer sheaths and sterile
coupling gel for clinical applications contacting compromised skin.
Some sheaths contain natural rubber latex and talc, which can cause
allergic reactions in some individuals.
Use market-cleared sheaths for clinical applications when Kosmos Torso
is likely to be splashed or splattered with blood or other bodily fluids.
Use market-cleared, sterile sheaths and sterile coupling gel to prevent
cross-contamination. Do not apply the sheath and coupling gel until you
are ready to perform the procedure. After use, remove and discard the
single-use sheath, and clean and disinfect Kosmos Torso using an
EchoNous-recommended high-level disinfectant.
After inserting Kosmos Torso into the sheath, inspect the sheath for holes
and tears.