KOSMOS Maintenance
72 KOSMOS User Guide
Kosmos Bridge
Cleaning
Avoid spraying the cleaning and disinfection solutions directly onto Kosmos
Bridge. Instead spray onto a non-abrasive cloth and then gently wipe. Ensure
that all excess solution is wiped off and not left on the surface after cleaning. The
following cleaning and disinfection method must be followed for
EchoNous
Kosmos Bridge Bridge.
1. After each use, disconnect the USB cable from Kosmos Torso.
2. Remove any accessories, such as the headset or power supply.
3. Using a wipe from an approved presaturated disinfectant wipe, carefully
wipe the screen and all other areas of Kosmos Bridge. Choose an EchoNous-
approved wipe from the list in Presaturated wipes.
4. If necessary, clean Kosmos Bridge with additional wipes to remove all visible
contaminants.
Kosmos Bridge is not sterile when shipped; do not attempt to sterilize it.
To avoid electrical shock, before cleaning, turn off Kosmos Bridge and
disconnect it from the power supply.
After disinfection, examine the display for cracks, and if damage
exists, discontinue use of the system and contact EchoNous
Customer Support.
TABLE 8-1. Presaturated wipes
Product Company Active Ingredients
Contact
Condition
Sani-Cloth
Plus
PDI Inc. n-Alkyl (68% C12, 32% C14)
dimethyl ethylbenzyl
ammonium chlorides. 0.125% n-
Alkyl (60% C14, 30% C16, 5%
C12, 5% C18) dimethyl benzyl
ammonium chlorides. 0.125%
5 minutes wet
contact time for
disinfection