KOSMOS User Guide 71
CHAPTER 8 KOSMOS Maintenance
Cleaning and disinfecting
General cautions
Some reprocessing chemicals may cause an allergic reaction in some
individuals.
Ensure that cleaning and disinfecting solutions and wipes are not
expired.
Do not allow cleaning solution or disinfectant into the Kosmos Bridge or
Kosmos Torso connectors.
Wear the appropriate personal protective equipment (PPE)
recommended by the chemical manufacturer, such as protective eye
wear and gloves.
Do not skip any steps or abbreviate the cleaning and disinfecting process
in any way.
Do not spray cleaners or disinfectants directly on Kosmos Bridge surfaces
or on Kosmos Bridge and Kosmos Torso connectors. Doing so may cause
solution to leak into KOSMOS, damaging it and voiding the warranty.
Do not attempt to clean or disinfect Kosmos Bridge, Kosmos Torso, or the
Kosmos Torso cable using a method that is not included here or chemical
not listed in this guide. Doing so can damage Kosmos Torso and void the
warranty.