KOSMOS Maintenance
74 KOSMOS User Guide
5. Using a new wipe, remove any particulate matter, gel, or fluids that remain on
Kosmos Torso using a new presaturated wipe from Presaturated wipes.
6. If necessary, clean Kosmos Torso with additional wipes to remove all visible
contaminants.
7. Before continuing with disinfection, ensure Kosmos Torso is visibly dry.
Disinfecting (intermediate-level)
Use the following steps to disinfect Kosmos Torso. Before performing the
following steps, read the following warnings and cautions.
To disinfect the Kosmos Torso (intermediate level):
1. After cleaning, choose an intermediate-level disinfectant from the list in
Presaturated wipes, and observe the recommended minimum wet contact
time.
2. With a new wipe, clean the cable and Kosmos Torso, starting from the
exposed cable, wiping toward the Kosmos Torso head to avoid cross-
contamination.
3. Observe the required wet contact time. Monitor Kosmos Torso for wet
appearance. Use at least three wipes to ensure effective disinfection.
4. Before reusing Kosmos Torso, ensure Kosmos Torso is visibly dry.
Always disconnect the USB cable from Kosmos Torso before cleaning and
disinfecting.
Always use protective eye wear and gloves when disinfecting any
equipment.
Before disinfecting, clean Kosmos Torso by following the appropriate
instructions to remove all gels, fluids, and particulates that may interfere
with the disinfection process.
Use only EchoNous-recommended disinfectants. Using a non-
recommended disinfecting wipe can damage Kosmos Torso and void the
warranty.
Check Kosmos Torso for damage, such as cracks, splitting, or sharp
edges. If damage is evident, discontinue using Kosmos Torso, and
contact your EchoNous representative.