Loading...
(ETS
FIBROSCAN
502
TOUCH
USER
MANUAL
FibroScan
Invalid
measurements
The
measurement
is
automatically
rejected
by
the
algorithms
if
the
pulse
sent
out
by the
transducer
could
not
be
delivered
successfully
and/or
if
the
shear
wave
propagation
maps
are
not
satisfactory.
The
message
'INVALID'
is
then
displayed
above
the
shear
wave
propagation
map.
6.6.6.
Shear
wave
speed
results
area
Median
Shear
wave
speed
is
expressed
in
meters
per
second
(m/s).
This
value
is
the
median
of
all
valid
measurements
performed
during
the
examination.
If
the
repeat
measurement
is
invalid,
the
median
is
not
recalculated.
To
obtain
a
reliable
and
representative
shear
wave
speed
measurement
in
the
liver,
at
least
10
valid
measurements
must
be
taken.
Refer
to
the
warning
in
Chapter
2
concerning
interpretation
of
the
result.
Interquartile
range
(IQR)
The
interquartile
range
(IQR)
is
expressed
in
meters
per
second
(m/s).
It
represents
the
interval
around
the
median
within
which
will
fall
50%
of
all
valid
measurements.
It
is
recalculated
after
each
new
valid
measurement.
36
03/2017
-
ECHOSENSTM
AND
FIBROSCAN®
ARE
TRADEMARKS
©
COPYRIGHT
ECHOSENS
-
ALL
RIGHTS
RESERVED
E300M017.4

Table of Contents