2. Measurement Preparations
Hint
Make sure that before each measurement
- the device has no visible defects
- the measuring system is tight (see leak test chapter 11.5)
2.1 Turn on the instrument and check the battery
The instrument is switched on with the <I / 0> key.
Hint
After switching on the instrument or before starting a measurement, check the
charging status of the battery.
Check the charging status of the battery in the status line and /
or in the menu "Control".
The instrument is supplied with charging power supply. The instrument can also be operated
with the internal battery for an extended period of time.
When recharging the battery is necessary, the instrument will emit a battery warning. This is
done acoustically and via a message in the display.
Hint
The battery warning is activated when the battery voltage is less than 6.5
V. Less than 6.0 V the operation with battery is no longer possible
The operation must continue via the charging power
supply.
WARNING
The instrument is operated with a Li-Ion battery!
The battery must be replaced only by qualified personnel or by ecom or an
ecom authorized service center.