SE-1 Electrocardiograph User Manual EMC Information
- 56 -
Appendix 2 EMC Information
Guidance and manufacture’s declaration – electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The SE-1 Electrocardiograph is intended for use in the electromagnetic environment specified
below. The customer or the user of the SE-1 Electrocardiograph should assure that it is used in
such an environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11 Group 1
The SE-1 Electrocardiograph uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class A
The SE-1 Electrocardiograph is suitable for
use in all establishments, other than domestic
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacture’s declaration – electromagnetic immunity –
For all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The Electrocardiograph is intended for use in the electromagnetic environment specified
below. The user of Electrocardiograph should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
It is recommended the use of
antistatic materials. If floor are
covered with synthetic material,
the relative humidity should be at
least 50%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for power
supply lines
It is recommended the use of
filters on power input lines and
enough separation between signal
lines and power lines.