4.5.2 Disposable Electrodes
Disposable Electrode Alligator Clip
Disposable electrode must be used together with the alligator clip.
The electrodes’ positions on body surface are as the following table and figures:
Above right ankle (Alternate placement, upper leg as close to torso
as possible)
Above left ankle (Alternate placement, upper leg as close to torso
as possible)
Fourth intercostals space at right border of sternum
Fourth intercostals space at left border of sternum
Fifth rib between V2 and V4
Fifth intercostals space on left midclavicular line
Left anterior axillary line at the horizontal level of V4
Left midaxillary line at the horizontal level of V4